157、selection and installation requirements of medical liquid oxygen storage tank (part 2) - doer equipment

classification:forum publishtime:2020-04-01 15:09:21

type selection and installation requirements of medical liquid oxygen storage tank (part 2)


the interval between liquid oxygen storage tank and pump house shall not be less than 3m, which is consistent with relevant foreign specifications and the actual practices of relevant domestic projects.


according to the analysis of the fire risk of medical liquid oxygen storage tank and its many years of operation experience, in order to meet the adjustment requirements of medical standards and the needs of hospital construction, the single tank volume and total volume of medical liquid oxygen storage tank are adjusted to 5m respectively ³ and 20m ³。


the fire prevention distance between medical liquid oxygen storage tank and buildings in medical and health institutions has been clearly specified in the national standard technical code for medical gas engineering (gb 50751-2012). the fire prevention distance between the medical liquid oxygen storage tank and the buildings outside the medical and health institutions shall still comply with the provisions of article 4.3.3 of this code.


article 4.6.4 of gb50751-2012 technical code for medical gas engineering also clearly stipulates the fire spacing of liquid oxygen storage tanks installed in hospitals;


the requirements for fire separation between medical liquid oxygen storage tanks and internal buildings and structures of medical and health institutions in gb50751-2012 technical code for medical gas engineering are as follows: